i-Checker - SERIES 170
SERIES 170The i-Checker is specially designed to calibrate measuringaccuracy of dial indicators, dial test indicators, and otherelectronic comparison gage heads with a stroke of up to 100mm(4").
Description
FEATURES
• ±(0.2+L/100)mm* indication accuracy
* When inspecting in vertical posture.
• Directly inspects an indicator with a stroke of up to 100mm
(4"). The dial test indicator, bore gage and lever-type inductive head can be inspected with optional accessories.
• Adjustment of the measurement position is very easily accomplished because of semi-automatic measurement and full automatic measurement functions.
• Creates and prints out the simple inspection certificate.
• Saves inspection result as CSV file for reusable inspection result by any kind of software.
SPECIFICATIONS | |
Model No. | i-Checker |
Order No. | 170-302■ (w ø8 bush) |
170-303■ (w ø3/8” bush) | |
170-302■-01 (w ø8 bush, calibration certificate) | |
170-303■-01 (w ø3/8" bush, calibration certificate) | |
Measuring Range | 100mm/4" |
Resolution | 0.02mm/1.24min |
Accuracy vertical position | ±(0.2+L/100)mm L = Arbitrary length |
(at 20°C) lateral position | ±(0.3+2L/100)mm L = Arbitrary length |
Drive method | Motor drive |
Measuring Unit | Reflective-type metal linear encoder |
Thermal expansion coefficient | (8±1)X10-6/K |
of measurement mode | |
Measurement method | Semiautomatic measurement |
Fully automatic measurement*1 | |
Digimatic Indicator with RS-232C output*2 | |
Dimensions (W x D x H) | 184 x 225 x 532mm (7.24" x 8.86" x 20.94") |
Operating temperature range | 20°C ±3°C |
Power supply | 100VAC to 240VAC ±10%, 50/60Hz |
Mass | 20kg/441lbs |
Note: To denote your AC line voltage fill the blank of order number with the following | |
suffixes (e.g. 170-302A). A for 120V, D and K for 220V, E for 240/220V. No suffix is | |
required for 100V. | |
*1Automatic measurement requires the indicator’s connection cable. Additionally some | |
form of indicator, along with a connecting machine (the optional accessory for indicator | |
as a Digimatic power-supply unit on EF counter) will be needed. | |
*2The indicator that can be measured via RS-232C is the one equipped with the function to | |
receive the “send data” request command form from the main unit and can output the | |
counter value to the main unit. |